Абстрактный

Radiation dose and image quality evaluation relative to different contrast media using cone-beam CT

Jijo Paul, Thomas J Vogl and Emmanuel C Mbalisike

Aims: To evaluate the radiation dose and image quality of an adult cone-beam CT (CBCT) with oil-based (OBC) and water-soluble contrast (WSC) material.
Material & Methods: A total of 44 patients, range: 46–82 years, (male:female – 26:18) who underwent CBCT examination during transarterial chemoembolization. Each patient received two CBCT scans; energy used: 94.4 ± 4.75 kV (mean ± standard deviation), current: 424 ± 132 mA for WSC and 94.5 ± 4.7 kV, 423.5 ± 132 mA for OBC. The volume of WSC material injected was 12 ml and OBC was 4 ml.
Result: WSC examination showed significantly (p < 0.05) decreased (5.83%) dose–area product compared with OBC. Hounsfield unit, noise, signal–noise ratio and contrast–noise ratio was higher for OBC (49.4, 19.44, 38 and 58%, respectively) compared with WSC (p < 0.05). Qualitative assessment of WSC data (median: 2, interquartile range: 1.5–2.5) showed higher image quality compared with OBC data (2.7, interquartile range: 2.3–3.9).
Conclusion: A detectable reduction of radiation dose was achieved with WSC compared with OBC in CBCT imaging. Quantitative image-quality parameters reflect higher values for OBC compared with WSC in the liver parenchyma. Subjective analysis showed an exactly opposite result due to the streak artifacts from OBC material.

Отказ от ответственности: Этот реферат был переведен с помощью инструментов искусственного интеллекта и еще не прошел проверку или верификацию